4-30 Programming Site Data
Entering Skan/Measure Count Parameters
To open the Skan/Measure Count window in the Skan/Measure
option, press the [X] arrow key twice (from the All tab) to reach
the Count tab. Press
[ENTER], and the screen appears similar to
Figure 4-13 below.
Figure 4-13: The Count Window in the
Skan/Measure Option
1. The first prompt asks for the Cycle Number that triggers the
count from 1 to 100. (The default number is 1.)
a. Press
[ENTER] to enter the text box.
b. Use the numeric keys to enter the recommended number.
c. Press
[ENTER] to confirm the entry.
2. The meter now asks for the Cycle Width (# of cycle count
samples, minimum width of valid cycle).
a. Press
[ENTER] to enter the text box.
b. Use the numeric keys to enter the recommended value.
c. Press
[ENTER] to confirm the entry.
Integrate
Skan/Measure
Cancell OK
All
Cycle Number
SNR Test
Cycle Width
Noise Level
%
Count
Correlate
1
Measure Only
Minimum Noise Level
8
20
5
%
Low SNR Error Level 7